Biomechanical Evaluation of Orthodontic Miniscrews: How Insertion Angle and Number of Insertions Impact Pull-Out Strength

Wilda Kemala Maulani*  -  Universitas Padjadjaran, Indonesia
Elih Sayuti  -  Universitas Padjadjaran, Indonesia
Endah Mardiati  -  Universitas Padjadjaran, Indonesia

(*) Corresponding Author

Background: This study aimed to analyze the effects of insertion angle and number of insertions on miniscrew pull-out strength in polyurethane bone blocks.

Method: In this experimental study, 30 miniscrews (Mico One; 2 mm diameter and 10 mm length) were inserted into Sawbones polyurethane blocks consisting of a 2 mm cortical layer (40 pcf) and an 18 mm trabecular layer (20 pcf). Samples were divided into six groups based on insertion angle (45°, 70°, and 90°) and number of insertions (single vs. repeated). Pull-out strength was measured using a Universal Testing Machine (UTM). Data were analyzed using two-way ANOVA followed by Tukey's HSD post hoc test, with significance set at p<0,05.

Result: In the single-insertion group, the insertion angle significantly affected pull-out strength (p<0,05). The 70° angle yielded the highest mean pull-out strength (339,98±20,76N), which was significantly higher than that of the 90° angle (299,93±20,45N; p<0,05). In the repeated-insertion group, the insertion angle did not significantly affect pull-out strength (p>0,05). Although the repeated-insertion group showed a lower mean pull-out strength (307,88±24,78  N) than the single-insertion group (318,34±25,21N), the difference was not significant (p>0,05).

Conclusion: Insertion angle significantly influences pull-out strength in single-insertion miniscrews, with 70° providing the highest strength. The number of insertions (single vs. repeated) does not significantly affect pull-out strength. These findings have implications for clinical anchorage strategies and experimental protocols using polyurethane bone analogs.

Keywords: Biomechanical phenomena, bone screw, materials testing, polyurethanes

  1. Ummat A, Shetty S, Desai A, Nambiar S, Natarajan S. Comparative assessment of the stability of buccal shelf mini-screws with and without pre-drilling- a split-mouth, randomized controlled trial. Clin Oral Investig. 2024;28(10):567. https://doi.org/10.1007/s00784-024-05925-7
  2. Antoszewska-Smith J, Sarul M, Łyczek J, Konopka T, Kawala B. Effectiveness of orthodontic miniscrew implants in anchorage reinforcement during en-masse retraction: A systematic review and meta-analysis. Am J Orthod Dentofac Orthop. 2017;151(3):440–55. https://doi.org/10.1016/j.ajodo.2016.08.029
  3. Bourassa C, Hosein YK, Pollmann SI, Galil K, Bohay RN, Holdsworth DW, et al. In-vitro comparison of different palatal sites for orthodontic miniscrew insertion: Effect of bone quality and quantity on primary stability. Am J Orthod Dentofac Orthop. 2018;154(6):809–19. https://doi.org/10.1016/j.ajodo.2018.02.010
  4. Jariyapongpaiboon P, ;Chartpitak J. Effect of Angular Insertion of Orthodontic Miniscrews on Primary Stability. Taiwan J Orthod. 2020;32(3). http://doi.org/10.38209/2708-2636.1032
  5. Lakshmikantha HT, Ravichandran NK, Jeon M, Kim J, Park HS. 3-Dimensional characterization of cortical bone microdamage following placement of orthodontic microimplants using Optical Coherence Tomography. Sci Rep. 2019;9(1):3242. http://doi.org/10.1038/s41598-019-39670-9
  6. Tepedino M, Masedu F, Chimenti C. Comparative evaluation of insertion torque and mechanical stability for self-tapping and self-drilling orthodontic miniscrews – an in vitro study. Head Face Med. 2017;13(1):10. http://doi.org/10.1186/s13005-017-0143-3
  7. Abidin SZ, Dasor MM. Methods for Assessing Orthodontic Mini Implant Stability. J Multidiscip Dent Res. 2024;10(2):68–73. https://doi.org/10.38138/JMDR/v10i2.18
  8. Nenen F, Garnica N, Rojas V, Oyonarte R. Comparison of the primary stability of orthodontic miniscrews after repeated insertion cycles: Angle Orthod. 2021;91(3):336–42. https://doi.org/10.2319/050120-375.1
  9. Araghbidikashani M, Golshah A, Nikkerdar N, Rezaei M. In-vitro impact of insertion angle on primary stability of miniscrews. 2016;150(3):436–443. http://dx.doi.org/10.1016/j.ajodo.2016.02.020
  10. Alkadhimi A, Al-Awadhi EA. Miniscrews for orthodontic anchorage: a review of available systems. J Orthod. 2018;45(2):102–14. https://doi.org/10.1080/14653125.2018.1443873
  11. Möhlhenrich SC, Heussen N, Modabber A, Bock A, Hölzle F, Wilmes B, et al. Influence of bone density, screw size and surgical procedure on orthodontic mini-implant placement – part B: implant stability. Int J Oral Maxillofac Surg. 2021;50(4):565–72. https://doi.org/10.1016/j.ijom.2020.07.003
  12. Wilmes B, Su Y yu, Drescher D. Insertion Angle Impact on Primary Stability of Orthodontic Mini-Implants. 2008;78(6):1065–70. http://dx.doi.org/10.2319/100707-484.1
  13. Alharbi F, Almuzian M, Bearn D. Miniscrews failure rate in orthodontics: systematic review and meta-analysis. Eur J Orthod [Internet]. 2018;40(5):519–30. https://doi.org/10.1093/ejo/cjx093
  14. Hong SB, Kusnoto B, Kim EJ, BeGole EA, Hwang HS, Lim HJ. Prognostic factors associated with the success rates of posterior orthodontic miniscrew implants: A subgroup meta-analysis. Korean J Orthod. 2016;46(2):111. http://synapse.koreamed.org/DOIx.php?id=10.4041/kjod.2016.46.2.111
  15. Baumgaertel S, Jones CL, Unal M. Miniscrew biomechanics: Guidelines for the use of rigid indirect anchorage mechanics. Am J Orthod Dentofac Orthop. 2017;152(3):413–9. http://dx.doi.org/10.1016/j.ajodo.2017.04.020
  16. Parinyachaiphun S, Petdachai S, Chuenchompoonut V. Considerations for placement of mandibular buccal shelf orthodontic anchoring screw in Class III hyperdivergent and normodivergent subjects – A cone beam computed tomography study. Orthod Waves. 2018;77(1):44–56. http://dx.doi.org/10.1016/j.odw.2018.01.001
  17. Xin Y, Wu Y, Chen C, Wang C, Zhao L. Miniscrews for orthodontic anchorage: analysis of risk factors correlated with the progressive susceptibility to failure. Am J Orthod Dentofac Orthop. 2022;162(4):e192–202. https://doi.org/10.1016/j.ajodo.2022.07.013
  18. Popa A, Dehelean C, Calniceanu H, Watz C, Brad S, Sinescu C, et al. materials A Custom-Made Orthodontic Mini-Implant — E ff ect of Insertion Angle and Cortical Bone Thickness on Stress Distribution with a Complex In Vitro and In Vivo Biosafety Profile. 2020;13(4789). https://doi.org/10.3390/ma13214789
  19. Nguyen M V., Codrington J, Fletcher L, Dreyer CW, Sampson WJ. The influence of miniscrew insertion torque. Eur J Orthod. 2018;40(1):37–44. https://doi.org/10.1093/ejo/cjx026
  20. Wang L, Oyonarte R, Carmona R, Bidart C, Battaglia G. Characteristics of the Buccal Shelf for the installation of miniscrews in Chilean individuals aged 15-45 years: a descriptive study. J Oral Res. 2024;13(1):222–33. https://doi.org/10.17126/joralres.2024.020
  21. Chhatwani S, Kouji-Diehl O, Kniha K, Modabber A, Hölzle F, Szalma J, et al. Significance of bone morphology and quality on the primary stability of orthodontic mini-implants: in vitro comparison between human bone substitute and artificial bone. J Orofac Orthop / Fortschritte der Kieferorthopädie. 2023;84(6):362–72. https://doi.org/10.1007/s00056-022-00385-8
  22. Aarash V, Abdollahi F, Ghorbanipour R, Mehdi M, Sistani N, Bijani A. Effect of insertion angles and reusing procedure on mini-screws’ mechanical properties. J Res Med Dent Sci. 2017;5(3):63–8. https://www.researchgate.net/profile/Ali-Bijani/publication/320279920_Effect_of_insertion_angles_and_reusing_procedure_on_mini-screws’_mechanical_properties/links/59dae2560f7e9b12b36d910b/Effect-of-insertion-angles-and-reusing-procedure-on-mini-screws-mec
  23. Meira TM, Tanaka OM, Ronsani MM, Maruo IT, Guariza-filho O, Camargo ES, et al. Insertion torque , pull-out strength and cortical bone thickness in contact with orthodontic mini-implants at different insertion angles. 2013;35:766–71. https://doi.org/10.1093/ejo/cjs095
  24. Mathews MC, PU B, Lakshmanan L, Joseph V. A Review on Mandibular Buccal Shelf Miniscrew. Cross Curr Int J Med Biosci. 2021;3(4):46–7. https://doi.org/10.36344/ccijmb.2021.v03i04.002
  25. Xu Z, Wu Y, Zhao L, Zhou Y, Wei X, Tang N, et al. Effect of placement angle on the stability of loaded titanium microscrews in beagle jaws. 2013;83(4). https://doi.org/10.2319/081612-660.1
  26. Kim GT, Jin J, Mangal U, Lee KJ, Kim KM, Choi SH, et al. Primary Stability of Orthodontic Titanium Miniscrews due to Cortical Bone Density and Re-Insertion. Materials (Basel). 2020;13(19):4433. https://doi.org/10.3390/ma13194433
  27. Hergel CA, Acar YB, Ateş M, Küçükkeleş N. In-vitro evaluation of the effects of insertion and sterilization procedures on the mechanical and surface characteristics of mini screws. Eur Oral Res. 2019;53(1):25–31. http://doi.org/10.26650/eor.20197993

Lisensi Creative Commons
This work is licensed under a Lisensi Creative Commons Atribusi-BerbagiSerupa 4.0 Internasional.
Contact us: Odonto Dental Journal: Jl. Raya Kaligawe Km.4, PO BOX 1054/SM Semarang, Central Java, Indonesia, 50112. Email: odontodentaljournal@unissula.ac.id
apps